CT/MRI
Á¦¸ñ | CBCT Phantom SedentexCT IQ |
Leeds Test Objects »çÀÇ CBCT ÆÒÅÒ SedentexCT IQ ÀÔ´Ï´Ù. A PMMA cylinder (160 mm diameter) with recesses to house test inserts (fig. 1). Within the body of the cylinder are features for the following tests; Noise/Uniformity The lower section of the phantom is uniform PMMA (density 1.20 +/- 1.00%) Geometric Distortion An array of 2.0 mm diameter, 3.0 mm deep Air gaps are uniformly pitched at 10.0mm intervals through one slice of the cylinder Test Inserts are included to perform the following measurements: Spatial Resolution Line Spread Function (LSF) - PMMA/PTFE interface (fig. 2) Point Spread Function (PSF) - 0.25mm diameter stainless steel wire suspended in air (fig. 3) LP/mm - alternating Aluminium/polymer (XY) (1.0, 1.7, 2.0, 2.5, 2.8, 4.0 and 5.0LP/mm) (fig.4) LP/mm - alternating Aluminium/polymer(Z) (1.0, 1.7, 2.0, 2.5, 2.8, 4.0 and 5.0LP/mm) (fig. 5) Contrast Resolution (fig. 6) 1.0, 2.0, 3.0, 4.0, 5.0 mm diameter Aluminium, PTFE, delrin, LDPE, Air, Water (PMMA) rods suspended in PMMA Pixel intensity (fig. 7) 10.0 mm diameter Aluminium, PTFE, delrin, LDPE, Air, Water (PMMA) rods suspended in PMMA Beam Hardening Artefacts (fig. 8) A line of three 5.0 mm diameter rods of Tisuspended in PMMA Blank PMMA insert (fig. 9) |
|
Ä«Å×°í¸® |
¹®ÀÇÇϱâ |