Product

´Ù¿ÃÆÒÅÒ Á¦Ç°À» ¼Ò°³ÇÕ´Ï´Ù.

Radiotherapy/Nuclear Medicine

Á¦¸ñ Pro-RT CTsim





This phantom is designed for quality control of a CT simulator, which usually consists of a CT scanner, virtual symulator software and a laser marking device (marking of the centre of target volume). Since geometrical planning is the core of CT simulation, periodic quality control is essential for maintaining optimum image quality and patient care.The phantom allows to perform detailed geometry and table movement tests as well as image quality assessment - low contrast resolution and high contrast detectability.

Technical data (can be modified to customer specifications):
overall dimensions: 150 x 150 x 150mm
made of PMMA
test object placed on four faces of the cube
Top Face contains contrast details objects (CD), MTF, ray line divergence (RLD) and spatial linearity (SL) patterns



First Side Face contains MTF and spatial linearity (SL) patterns

Second and Third Side Face contains ray line divergence (RLD) pattern

optional carrying case
Complies with: – IEC 61223-3-5 – IEC 61223-2-6
CE certified
manual provides detailed guidelines for carrying out each test, results assessment and registration

Product features:
• Complies with:
– IEC 61223-3-5
– IEC 61223-2-6
• CE certified
• manual provides detailed guidelines for carrying out each test, results assessment and registration

¹®ÀÇÇϱâ